> 제품소개 > 상세설명
특징 (Features)
USB
RS-232
GPIB
Handler
SCAN
주요 특징
50,000 카운트
디스플레이
3.5인치(320x240) TFT LCD 디스플레이 0.05% 높은 정확도 1A 테스트 전류, 1uΩ 분해능 측정 속도 : 최대 60
read/s 4선 측정
방식 온도 보정 측정
기능 지연 측정
기능 20세트 패널 설정
메모리 Dry 회로
(GOM-805) 드라이브
모드 : DC+/DC-, Pulsed, PWM, Zero(GOM-805) 기본 인터페이스 : USB, RS-232C,
Handler/Scan/EXT I/O 옵션 인터페이스 : GPIB (GOM-805는 기본
장착)
GW Instek launch a new series of D.C. milliohm meter ─ GOM-804/805, which
abundantly feature 3.5-inch TFT display, maximum 50,000 counts measurement
display, the rapid sampling rate of 60 readings per second, optimum 0.05%
measurement precision, four wire measurement method as well as the temperature
measurement and temperature compensation measurement function to meet the
requirement of low resistance measurement application. The GOM-805 also includes
various drive modes and Dry circuit for contact resistance measurement
applications. More features, including 20 sets of panel setting memory and many
external control interface such as RS-232C, USB, Handler/Scan/EXT IO or GPIB
(option), greatly elevate GOM-804/805 milliohm meter’s convenience on practical
applications.
GOM-804/805 adopt 3.5-inch color LCD to enhance the clarity of measurement
results and to provide display for related setting criteria that tremendously
brings up the completeness of test information. Additionally, GOM-804/805, with
the optimum 0.05% precision, augment the measurement speed to 60 sampling rate
per second and maintain the display digits of five instead of four despite of
different speed selections. Furthermore, the independent functionality keys and
direction keys together increase the operational convenience which allows users
to complete their measurement tasks with intuitive convenience and speed.
GOM-805 provides Dry circuit and various drive modes (DC+, DC-, Pulsed, PWM)
for measurement applications on different materials. The pulsed current output
mode is suitable for interacting conductors of different materials and this
output mode is to reduce the thermal EMF influence, which is caused by electric
potential difference generated from different conductors acting on different
temperatures while conducting low resistance measurements. The DC+ and DC-
output modes are best for the measurement requirements of inductive components.
The PWM output mode, ideal for changing temperature sensitive materials, can
avoid resistance value variation which is due to over load happened on current
measurement for a long period of time. During the DC+, DC- and Pulsed drive is
supplied; the Dry circuit can work with them also. Dry circuit can limit the
applied voltage under the open circuit voltage of 20mV to avoid over voltage
occurred on the both ends of components. The over voltage will damage the oxide
coating and the thin layer of contact surface, as a result, the validity of
measurement will then be ruined. For instance, contact resistance of connector
measurement is one of the applications.
With respect to connecting the external control, GOM-804/805 provide a D-sub
25-pin combined interface to execute, according to the functionalities, Handler,
Scan or EXT IO for respectively connecting to a sorting machine; connecting to
an external on-off switch, and directly conducting external trigger control. For
remote control and measurement result retrieval requirements, GOM-804/805 also
provide various interface selections such as RS-232C, USB, and GPIB
(GOM-804(option)/GOM-805(standard) interface. Furthermore, the control commands
are compatible to that of GOM-802 that saves time in adjusting programs while
switching from the old model to the new model.
To sum up, GOM-804 evolves from GOM-802 platform with more advanced
functionalities and specifications, including display digits, measurement speed
and standard interface (RS-232C/USB). With all the capabilities of GOM-804,
GOM-805 augments itself with new measurement abilities (Dry circuit and various
drive modes) to meet the requirements of broader low resistance measurement
applications.
Totally replacing the existing models
In terms of the basic functionalities and specifications,
GOM-804/805 can absolutely replace the existing model─GOM-802. All GOM-802
functionalities can be found from GOM-804/805, including resistance measurement
range, 1A test current (maximum), four wire measurement method, temperature
probe (option, accessory model: PT-100) for temperature measurement and
temperature compensation measurement, etc. The programming commands are also
compatible to that of GOM-802. To simply put it, the brand new GOM-804/805 not
only provide better display interface, fast measurement (60 readings per
second), but also collocate with standard communications interface (RS-232C/USB
device) to facilitate users in accomplishing measurement tasks rapidly. On top
of that, model switching will not be a problem.
Faster measurement without sacrificing
resolution
GOM-804/805 has two measurement speed selections, which are Fast
reaching 60 readings per second, and Slow 10 readings per second. A major
departure from the past, users, in the past, had to juggle between speed and
display resolution. GOM-804/805 will not affect resolution despite of any speed
selections and will maintain the highest display digits. In other words, reading
resolution will not be changed by changing speed and the display digits remain
the same.
Dry circuit test for GOM-805 only
Dry circuit is to limit test voltage and current to certain levels which will
not cause contact points to produce physically or electrically changed circuit
and its most frequently used application is contact resistance of connector
measurement. Based upon MIL-STD-1344 method 3002-1 low signal level contact
resistance, tests must be applied under the maximum open circuit voltage of 20mV
(or lower), and short circuit current of 100mA (or lower) to avoid over voltage
for the both ends of components. The over voltage will damage the oxide coating
and the thin layer of contact surface, as a result, the validity of measurement
will then be ruined. GOM-805 provides three levels (500mΩ:100mA/5Ω:10mA/50Ω:1mA)
to limit open circuit voltage at 20mV to execute Dry circuit
tests.
Various drive modes for GOM-805 only
GOM-805 provides various current output drive modes to satisfy diversified
and accurate low resistance measurement applications. For instance, for
interacting conductors of different materials, the pulsed current output mode
can be applied to reduce the thermal EMF influence, which is caused by different
conductors acting on different temperatures. The PWM output mode, ideal for
changing temperature sensitive materials, can avoid resistance value variation
which is due to over load on large current measurement in a long period of time.
The DC+ and DC- output modes are best for the measurement requirements of
inductive components.
Standard interface for control and communications
With respect to connecting the external control, GOM-804/805 provide a D-sub
25-pin composite interface to execute, according to the functionalities,
Handler, Scan or EXT IO for connecting to a sorting machine; connecting to an
external on-off switch, and directly conducting external trigger control
respectively. For remote control and measurement result retrieval requirements,
GOM-804/805 also provide various interface selections such as RS-232C, USB, and
GPIB (GOM-804(option)/GOM-805(standard) interface.
Command compatiability
The commands of GOM-804/805 are compatible to that of GOM-802 that allows
users to switch equipment with simple settings. There is no cost in adjusting
existing programs and production delay will not be happening while switching
from the old model to the new model.
GOM-804/805
DISPLAY
3.5"(320x240) TFT LCD, 50000 Counts
Sampling Rate
Slow
10 readings/s
Fast
60 readings/s
RESISTANCE
Range/Test Current
50mΩ/1A, 500mΩ/100mA, 5Ω/100mA, 50Ω/10mA, 500Ω/1mA, 5kΩ/100uA, 50kΩ/100uA,
500kΩ/10uA, 5MΩ/1uA 9 ranges
Accuracy
50mΩ : ±(0.1% of reading + 0.02% of range)500mΩ ~ 50Ω : ±(0.05% of
reading + 0.02% of range)500Ω ~ 500kΩ : ±(0.05% of reading + 0.008% of
range)5MΩ : ±(0.2% of reading + 0.008% of range)
Resolution
1uΩ, 10uΩ, 100uΩ, 1mΩ, 10mΩ, 100mΩ, 1Ω, 10Ω, 100Ω
TEMPERATURE
Range
-50°C ~ 399.9°C
Accuracy
-10°C ~ -40°C : 0.3%+0.5°COther : 0.3%+1.0°C
Resolution
0.1°C
DRY CIRCUIT (GOM-805 Only)
Open circuit less than 20mV; For 500mΩ, 5Ω, 50Ω range only
DRIVE MODE (GOM-805 Only)
DC+ / DC-
Yes
Pulsed
Yes
PWM
Yes
Zero
Yes
FUNCTION
Trigger ? Internal, Manual, External、Math ? ABS, REL, %, TC、Hi / Lo, Average
2 ~ 10 times、Measured delay、Go / No-Go、TC for Transformer, Diode、Continuity
beeper、Sorting (only for GOM-805)
MEMORY
20 Sets
INTERFACE
Standard : USB, RS-232C, Handler/Scan/EXT IOOption : GPIB (Standard for
GOM-805)
POWER SOURCE
AC 100V ~ 240V, 50/60HzPower Consumption : Max 25VA
DIMENSIONS & WEIGHT
223(W)×102(H)×283(D) mm, Approx. 4kg